Hi friends!!!
In our previous blogs, we have discussed about the following topics.
EU MDR Timeline, Annex, Chapter
EU MDR Post Market Surveillance
EU MDR Technical Documentation - Part 1
EU MDR Technical Documentation - Part 2
EU MDR General Safety and Performance Requirement
EU MDR Information on the Label
EU MDR Post Market Clinical Follow Up
EU MDR Unique Device Identification - Part 1
EU MDR Unique Device Identification - Part 2
EU MDR Chapter & Articles - Part 1
EU MDR Chapter & Articles - Part 2
EU MDR 2017/745 is live now. And it is the most trending topic in Medical Device Regulatory Affairs field. Most of the medical device companies are in EU MDR Transition to make sure that they can sell their products in Europe.
Therefore, it is important for Medical Device QA/RA Professionals to know in detail about EU MDR 2017/745. In today's blog, we are going to discuss about Scrutiny Procedure of EU MDR 2017/745.
Let's get into the topic.
SCRUTINY PROCEDURE
The Scrutiny procedure in EU MDR is intended to
improve the overall quality of conformity assessments performed by the notified
body for high-risk category devices. The Scrutiny is for additional review
& improves transparency in the conformity assessment process.
As per Article 54 & 55, manufacturers of
Class III Implantable Devices & Class IIb active devices intended to
administer or remove medicinal products are subjected to scrutiny.
Apart from Manufacturers of Medical Devices, Scrutiny also impacts the Notified Bodies, Competent Authority, MDCG & Expert Panel.
Image courtesy: WikipediaLicensed under: CC Share Alike
The following are the roles &
responsibilities of Manufacturers, Notified Body, Competent Authority, MDCG
& Expert Panel:
Manufacturer:
The Manufacturer of Medical Device submits the
Conformity Assessment documentation & Summary of Safety & Clinical
Performance to the Notified Body.
Notified Body:
The Notified Body performs conformity
assessment of the documentation submitted by the manufacturers of the Medical
Device. As per “Article 55, Mechanism for scrutiny of conformity assessments of
certain Class III & Class IIb devices”, the Notified Body notifies the
competent authority about the certificates it has granted by performing
conformity assessment in accordance with Article 54(1). Such notification shall
take place via electronic system as referred in Article 57. The notification
shall include the following details.
The Summary of Safety and Clinical Performance
in accordance with Article 32.
Conformity Assessment report by the notified
body.
The Instructions for Use as referred in Section
23.4 of Annex I.
If applicable, scientific opinion of the expert
panels as referred in Section 5.1 of Annex IX or Section 6 of Annex X. If the
Notified Body decided not to follow the advice from Expert Panel, an
appropriate justification shall also be included.
Competent Authority:
The Competent Authority determines the need to
perform the additional scrutiny of the Notified Bodies Conformity Assessment of
Documentation. The Competent Authority makes such decisions in accordance with
the following articles.
Article 44: Monitoring & re-assessment of
notified bodies
Article 45: Review of notified body assessment of
technical documentation and clinical evaluation documentation
Article 46: Change to designations &
notifications
Article 47: Challenge to the competence of
notified bodies
Article 94: Evaluation of devices suspected of
presenting an unacceptable risk or other non-compliance
The Competent Authority also coordinates with
MDCG, if required.
Medical Device Coordination group (MDCG):
The MDCG directs the scrutiny process &
coordinates with the expert panel, if required.
Expert Panel:
The Expert Panel provides scientific expertise
& expert opinion on the conformity assessment. The Expert Panel can advise
restrictions to medical indications or intended purpose, can limit the
certificate validity, propose updates to PMCF Plans, IFU’s, SSCP.
Article 54: Clinical Evaluation consultation procedure for certain
class III and class IIb devices
When performing a conformity assessment of
Class III & Class IIb Medical Devices, apart from the procedures specified
in “Article 52, Conformity Assessment Procedures”, the notified body can also
follow the procedure regarding Clinical Evaluation Consultation as specified in
Section 5.1 of Annex IX or Section 6 of Annex X.
The notified body shall notify the competent
authority via electronic system as referred in Article 57 of whether the above-mentioned
procedure of article 54 is applied or not. The notification shall be
accompanied by clinical evaluation assessment report.
Now I hope you have acquired some basic knowledge about Scrutiny Procedure as per EU MDR 2017/745. We will meet again with another valuable topic related to Biomedical Engineering, Medical Device, Healthcare.
NOTE: Dear friends!!!... Please do comment a topic related to Biomedical, so that we can discuss it in future blogs.
Check out for this blog about Computed Tomography, X ray, MRI scan, ECG.
COMPUTED TOMOGRAPHY|TECHNOLOGY|ARTIFACTS
x ray machine working principle
Check out for the below blogs about IEC 62304 & IEC 60601-1.
IEC 62304 Software Configuration Management & Problem Resolution Process
IEC 62304 Software Maintenance Process
IEC 62304 Software System Testing
IEC 62304 Software Unit Implementation & Integration
IEC 62304 Software Architectural & Detailed Design
IEC 62304 Software Requirements Analysis
IEC 62304 Software Development Planning
IEC 62304 Overview & Definitions
IEC 60601-1 Classification of Medical Electrical Equipment
IEC 60601-1 General Requirements for Testing Medical Electrical Equipment
IEC 60601-1 Medical Electrical Systems
IEC 60601-1 Protection against unwanted and excessive radiation hazards
IEC 60601-1 General Requirements
IEC 60601-1 Programmable Electrical Medical Systems
IEC 60601-1 Hazardous situations and fault condition
IEC 60601-1 Construction of Medical Electrical Equipment
IEC 60601-1 Leakage Currents and Patient Auxiliary Currents
IEC 60601-1 Creepage Distance and Air Clearance
IEC 60601-1 Components and Wiring
IEC 60601-1 Mechanical Hazards Moving Parts
IEC 60601-1 Excessive Temperatures
IEC 60601-1 Markings on the Outside of Medical Electrical Equipment
IEC 60601-1 Markings on the Inside of Medical Electrical Equipment
IEC 60601-1 Instability Hazards
IEC 60601-1 Separation of Parts
DON’T FORGET TO FOLLOW THIS BLOG PAGE
DROP YOUR MAIL ID TO GET IMMEDIATE UPDATES
If you find this blog post knowledgeable,
comment & share it with a friend!