In today’s blog, we are going to discuss about IEC 60601-1.
IEC 60601-1 states the Basic safety requirements for Medical Electrical Equipment’s. Satisfying the requirements of IEC 60601-1 is an important task in order to commercialize the Medical Electrical Equipment’s
Therefore, it is important for Medical Device QA/RA Professionals to know in detail about Clause 14: Programmable Electrical Medical Systems as per IEC 60601-1.
Let’s get into the topic.
Clause 14 – Programmable
Electrical Medical Systems
The Clause 14 of IEC 60601-1
states the requirements of Programmable Electrical Medical Systems. The
following are the Sub-Clauses.
General
Documentation
Risk Management plan
PEMS Development Life-Cycle
Problem resolution
Risk Management Process
Requirement specification
Architecture
Design and implementation
Verification
PEMS Validation
Modification
PEMS intended to be incorporated
into an IT-Network
Let’s discuss in detail about
Important requirements of above-mentioned Sub Clauses.
Risk Management Plan:
The PEMS Validation Plan shall be
referenced in Risk Management Plan.
PEMS Development Life-Cycle:
Milestones shall be defined at
each stage of PEMS Development Life-Cycle. Each milestone shall have the
details of verification methods and activities to be completed. Each activity
listed in the milestone shall have details about its input and output. The PEMS
Development Life-Cycle shall follow risk-based approach.
The PEMS Development Life-Cycle
has certain attributes. All the activities in PEMS Life-Cycle shall be grouped
into decomposition process and integration process.
Requirement Specification:
Each Subsystem of PEMS shall have
requirement specification document. The requirement specification shall have
details about essential performance of Subsystem and risk controls implemented.
Architecture:
Each Subsystem of PEMS shall have
architecture which satisfies the requirement specification.
The architecture shall be framed
to reduce the risk to an acceptable level by making use of Components with High
Integrity, fail safe functions, redundancy, defensive design, test interval
duration & diagnostic coverage, protection from reasonably foreseeable
misuse, etc.
Design and Implementation:
Each Subsystems of PEMS shall
have a Design and Test Specification.
Verification:
Each function which ensures basic
safety, essential performance, and risk control measures shall be verified. The
Verification shall be performed for each function at each milestone. A
Verification Plan shall be created to decide on the strategies and methods of
verification. The strategies and methods of verification shall include
Statistical Techniques, Inspections, Dynamic Analysis etc.
The Verification shall be
performed as per the plan and the results shall be documented.
Validation:
The Person responsible for
overall validation shall be independent from the Design Team. No member of the
design team shall perform PEMS Validation of their own design. The rationale
for the level of independence shall be given by the manufacturer.
The Validation shall be performed
as per the plan and the results shall be documented.
Programmable Electrical Medical Systems intended to be incorporated into an IT-Network:
If the PEMS is intended to be
incorporated into an IT-Network which is not validated by Manufacturer. The instructions
for implementing such connection shall be made available by the manufacturer
which shall include the following:
The purpose of the Programmable
Electrical Medical Systems connection to an IT-Network shall be available.
The required characteristics and
configuration of the IT-Network incorporating the PEMS.
The technical and security
specifications of the network shall be available.
During the incorporation of PEMS
to IT Network, there will be subsequent changes in the IT Network which can
introduce new risks and raises the requirement to perform additional analysis.
The changes to the IT Network
shall include the following:
The change in the configuration
of IT Network.
Disconnection of items from the
IT Network.
The update of equipment connected
to the IT Network
Now I hope you have acquired some knowledge about IEC 60601-1 Clause 14: Programmable Electrical Medical Systems. We will meet again with another valuable topic related to Biomedical Engineering, Medical Device, Healthcare.
NOTE: Dear friends!!!… Please do comment a topic related to Biomedical, so that we can discuss it in future blogs.