Hi friends!!!
In our previous blogs, we have discussed about the following topics.
IEC 60601-1 Classification of Medical Electrical Equipment
IEC 60601-1 General Requirements for Testing Medical Electrical Equipment
IEC 60601-1 Medical Electrical Systems
IEC 60601-1 Protection against unwanted and excessive radiation hazards
IEC 60601-1 General Requirements
IEC 60601-1 Programmable Electrical Medical Systems
IEC 60601-1 Hazardous situations and fault condition
IEC 60601-1 Construction of Medical Electrical Equipment
IEC 60601-1 Leakage Currents and Patient Auxiliary Currents
IEC 60601-1 Creepage Distance and Air Clearance
IEC 60601-1 Components and Wiring
IEC 60601-1 Mechanical Hazards Moving Parts
IEC 60601-1 Excessive Temperatures
IEC 60601-1 Markings on the Outside of Medical Electrical Equipment
EU MDR Timeline, Annex, Chapter
EU MDR Post Market Surveillance
EU MDR Technical Documentation – Part 1
EU MDR Technical Documentation – Part 2
EU MDR General Safety and Performance Requirement
EU MDR Information on the Label
EU MDR Post Market Clinical Follow Up
EU MDR Unique Device Identification – Part 1
EU MDR Unique Device Identification – Part 2
EU MDR Chapter & Articles – Part 1
EU MDR Chapter & Articles – Part 2
In today’s blog, we are going to discuss about IEC 60601-1.
IEC 60601-1 states the Basic safety requirements for Medical Electrical Equipment’s. Satisfying the requirements of IEC 60601-1 is an important task in order to commercialize the Medical Electrical Equipment’s
Therefore, it is important for Medical Device QA/RA Professionals to know in detail about Sub Clause 7.3: Markings on the Inside of Medical Electrical Equipment as per IEC 60601-1.
Let’s get into the topic.
Markings on the Inside of Medical Electrical Equipment
Heating Elements or Lamp Holders:
The Heating elements or lamp holders designed to be used with heating lamps shall be marked with information about maximum power loading near the heater or in the heater itself.
High Voltage Parts:
The Medical Electrical Equipment’s parts of high voltage shall be marked with the symbols of “Dangerous Voltage”, “Warning”, “Electricity” as specified in IEC 60601-1.
Batteries:
The markings for Type of Battery and Mode of Insertion shall be provided in the Medical Electrical Equipment.
In case the battery can only be removed by the service personnel with the use of tool, the marking referring to the information stated in the accompanying documents is sufficient. If any hazardous situation can happen due to incorrect replacement of battery like reversed polarity, a warning shall be provided along with the marking stating that replacement by inadequately trained personnel could lead to hazardous situation.
Fuses, Thermal Cuts, Over Current Releases:
Fuses, Replaceable Thermal Cuts, Over Current Releases of the Medical Electrical Equipment which can only be removed with the use of a tool shall be marked with the specifications adjacent to the component or shall be referred to the information in the accompanying documents.
Protective Earth Terminal:
Unless the Protective Earth Terminal is an appliance inlet, it shall be marked with the following symbol as specified in IEC 60601-1 .
Functional Earth Terminal:
The Functional Earth Terminal of Medical Electrical Equipment shall be marked with the following symbol as specified in IEC 60601-1
Supply Terminals:
The Terminals which are supposed to be connected to the supply conductors shall be marked adjacent to the terminals.
The Terminals which are supposed to be connected exclusively to the neutral supply conductors shall be marked with the code as specified in IEC 60601-1,
Temperature of Supply Terminals:
If the Terminal Box or wiring compartment attains a temperature of more than 75 °C during normal use and normal condition when it is intended to be used in connection with power supply conductors for permanently installed Medical Electrical Equipment, in this case the Medical Electrical Equipment shall be marked with information related to the minimum temperature requirements of wiring materials to be used in supply connections.
This information shall be marked near to or at the place where supply connections are to be made.
Markings of Controls and Instruments:
Power Switches:
The Switches which are used to Control the power of the Medical Electrical Equipment or its parts shall have its ON and OFF positions indicated by adjacent light & marked with symbols as specified in IEC 60601-1.
Control Devices:
Figures, letters or other visual means shall be used to indicate the different positions of Control Devices and different positions of switches in the Medical Electrical Equipment.
In case the change of settings in a control during normal use shall lead to an unacceptable risk, such controls shall be provided with an indicating device.
Unit of Measurement:
Numeric Indications of parameters in the Medical Electrical Equipment shall be expressed in SI Units.
Now I hope you have acquired some knowledge about IEC 60601-1 Sub Clause 7.3: Markings on the inside of Medical Electrical Equipment. We will meet again with another valuable topic related to Biomedical Engineering, Medical Device, Healthcare.
NOTE: Dear friends!!!… Please do comment a topic related to Biomedical, so that we can discuss it in future blogs.